Test Code OPATM OPIATE CONFIRMATION, MECONIUM

Specimen

Meconium

Collect meconium specimen and submit in a screw capped container or stool container. Make sure to collect the entire random meconium specimen.

 

Specimen stability:

14 days at room temperature

28 days refrigerated at 2-8°C

28 days frozen at ≤ -20°C  (preferred)

Minimum volume:

0.3 g (approximately 1 teaspoon)

Specimen rejection criteria:

Grossly bloody specimens, pink is OK

Stool

Diapers
 

Reference Values

Negative

 

Cutoff concentrations for LC-MS/MS testing:

Codeine: 20 ng/g

Hydrocodone: 20 ng/g

Hydromorphone: 20 ng/g

Morphine: 20 ng/g

Oxycodone: 20 ng/g

Oxymorphone: 20 ng/g

Day(s) Test Performed

Monday - Sunday

Performing Laboratory

Mayo Clinic Laboratory - Rochester

CPT Coding

80361

80365

G4080 (if appropriate)

Useful For